Home |
Current Issue |
Past Issues |
In the Clinic |
ACP Journal Club |
CME |
Collections |
Audio/Video |
Mobile |
Subscribe |
Tools |
Help |
ACP Online
|
1 April 1994 | Volume 120 Issue 7 | Pages 552-558
Objective: To enhance understanding of the reliability of the international normalized ratio (INR) for monitoring warfarin therapy and its relation to other monitoring techniques.
Design: Prospective cohort study.
Setting: A university hospital.
Patients: 79 patients attending an anticoagulation clinic.
Measurements: International normalized ratios obtained with a portable capillary monitor (Coumatrak) and the following from a simultaneous plasma sample: INRs from prothrombin times done with six thromboplastins, prothrombin-proconvertin (P&P) test activity, specific prothrombin activity, and native prothrombin antigen.
Results: Converting to INRs failed to standardize prothrombin time results obtained with high- and low-sensitivity thromboplastins. Coumatrak INRs correlated best with INRs obtained with high-sensitivity thromboplastins. The INR range of 2.0 to 3.0 corresponded to a P&P range of 30% to 13%, a native plasma prothrombin antigen range of 56 to 24 µg/mL, and a specific prothrombin activity range of 43% to 21%.
Conclusions: Low-sensitivity thromboplastins may give erroneously high INRs in the upper therapeutic range. Plasma prothrombin times should be done with a high-sensitivity thromboplastin, particularly in patients maintained at the upper limit of the therapeutic range. An INR so obtained correlated well with an INR obtained with a portable capillary blood monitor.
In 1992, a prothrombin time ratio was still used to monitor warfarin therapy at our medical center. The ratio was determined for inpatients from a plasma prothrombin time and for outpatients from a capillary whole-blood prothrombin time measurement system (the Coumatrak portable monitor [Du Pont Pharmaceuticals; Wilmington, Delaware]). To prepare for reporting prothrombin time results as INRs, we compared the INRs as routinely determined by the capillary monitor in a large group of outpatients on maintenance warfarin therapy with the INRs as determined by plasma prothrombin times done with six commonly used thromboplastins. We also measured the prothrombin time by the so-called prothrombin-proconvertin (P&P) method [6], residual plasma native prothrombin antigen [7], and specific prothrombin (factor II) activity [8].
This protocol was adopted after a preliminary experiment was done in which blood samples from five patients were divided into two aliquots and allowed to stand for either 2 hours or 5 hours before separating the plasma and measuring the prothrombin time. The prothrombin time values, as measured with three different thromboplastins, did not differ significantly for the paired samples (data not shown). In addition, virtually identical P&P test results were obtained on fresh plasma samples from five patients and on these samples after they had been stored frozen for 1 week.
This test was done by two experienced practitioners on finger-stick blood. The thromboplastin used in the cartridge for the Coumatrak instrument had an ISI of 2.04. The value for a normal prothrombin time was set by the instrument at 12 s, and the prothrombin time ratio and INR were provided by the instrument as a direct readout.
Prothrombin time ratios for assays done with each thromboplastin were calculated using a normal value obtained with each thromboplastin for each run with the control pooled plasma. The mean clotting times (±SD, n = 7) of the control pooled plasma with the different thromboplastins were as follows: Dade C, 11.8 ±0.2 s; Dade C-plus, 12.2 ±0.2 s; Dade IS, 13.7 ±0.2 s; Excel, 11.7 ±0.1 s; Excel-S, 13.4 ±0.2 s; and Ortho, 12.7 ±0.3 s. The INR was calculated as INR = PTRISI with the manufacturer's value for the photo-optical ISI as given in the package insert.
The thromboplastins used in this study had ISI values ranging from 1.27 to 2.88 (see Methods). Dade Thromboplastin IS (Dade IS; ISI, 1.27) was the most sensitive of the thromboplastins and was therefore selected as the standard against which we compared data obtained with the other thromboplastins.
In an initial analysis, the prothrombin time values obtained with the Coumatrak capillary monitor and with Dade IS were plotted both as prothrombin time ratios Figure 1 A and as INR values Figure 1 B. Because of the different ISI values for the Coumatrak monitor and Dade IS, the slope of the linear regression line for the data plotted as prothrombin time ratios, 0.36 (r = 0.87, n = 100), clearly differed from 1. The slope of the regression line of the data plotted as INRs, 0.84 (r = 0.90, n = 100), was closer to 1, which is the theoretically expected slope when plotting INR values against each other.
ARTICLE
The International Normalized Ratio (INR) for Monitoring Warfarin Therapy: Reliability and Relation to Other Monitoring Methods
Reporting prothrombin times used to monitor warfarin therapy as an international normalized ratio (INR) has been proposed as a way to eliminate interlaboratory differences in test results caused by the use of thromboplastins with different sensitivities [1-4]. The INR is calculated by raising the prothrombin time ratio (PTR; the patient's prothrombin time divided by a reference normal prothrombin time) to the power of a coefficient known as the international sensitivity index (ISI). This coefficient relates the sensitivity for monitoring warfarin therapy of a given thromboplastin to the sensitivity of the World Health Organization's first primary international reference preparation of thromboplastin, which was assigned an ISI of 1.0 [5]. Thromboplastins less sensitive than this international reference preparation of thromboplastin to decreased levels of the vitamin K-dependent clotting factors have proportionately higher ISI values.
Methods
![]()
Top
Methods
Results
Discussion
Author & Article Info
References
One hundred plasma samples were collected from 79 patients between 8 April and 20 May 1992 at seven weekly sessions of the outpatient anticoagulant clinic of the University of California, San Diego, Medical Center. All samples except one were obtained from patients who had taken warfarin for longer than 14 days. All patients gave informed consent. After the Coumatrak prothrombin time was measured, 4.5 mL of venous blood was collected in a siliconized BD Vacutainer tube (Becton-Dickinson; Rutherford, New Jersey) containing 0.5 mL of 3.2% sodium citrate and sent to the medical center's Special Coagulation Laboratory. All subsequent handling and testing were done by the same experienced medical technologist who followed the routine techniques used in this laboratory. Platelet-poor plasma was prepared within 2 hours of the time of sampling by centrifugation at 3000 rpm for 15 minutes at room temperature. In the early afternoon, when all samples for a given day had been processed, prothrombin times were done on each sample with six thromboplastin reagents. The plasma samples were then stored at 70°C for subsequent measurement of the P&P percent activity, specific prothrombin (factor II) activity, and native prothrombin antigen.
Control Pooled Plasma
![]()
The normal plasma used to obtain the prothrombin time ratio for plasma prothrombin times was the routine control pooled plasma of the Special Coagulation Laboratory. This plasma is prepared by pooling plasma obtained by the technique described above from blood collected in 1/10 volume of a buffered citrate anticoagulant (0.06 mol/L sodium citrate plus 0.04 mol/L citric acid) from between 15 and 20 healthy donors. It is stored in small aliquots at 70°C. The prothrombin time and activated partial thromboplastin time of the plasma from each donor had to be within 2 standard deviations (SD) of the mean of the individual plasma samples from the donors that were included in the laboratory's previous batch of control pooled plasma.
Coagulation Assays
![]()
Capillary Blood Prothrombin Time (Coumatrak)
Prothrombin Time Assays
![]()
The prothrombin times were done on an Electra 800 (Medical Laboratory Automation; Pleasantville, New York). The six thromboplastins, made available by sales representatives of the manufacturers, were Thromboplastin C (Dade C; ISI, 2.88), Thromboplastin C-plus (Dade C-plus; ISI, 2.03), and Thromboplastin IS (Dade IS; ISI, 1.27) from Dade division, Baxter Scientific Products, Miami, Florida; Simplastin Excel (Excel; ISI, 1.99 or 2.04 for two lots used) and Simplastin Excel-S (Excel-S; ISI, 1.34) from Organon Teknica, Durham, North Carolina; and Ortho Brain Thromboplastin (Ortho; ISI, 2.30) from Ortho, Raritan, New Jersey.
Prothrombin-Proconvertin Test
![]()
This test was done according to the manufacturer's instructions with Simplastin A (Organon Teknika), which is the source not only of the thromboplastin (ISI, 1.70) and CaCl2 but of an optimal concentration of factor V and fibrinogen. The test plasma was diluted in saline 1:10, which sensitized the assay to small changes in the prothrombin, factor VII, and factor X activity of the sample. Clotting times were determined with an ST-4 semi-automated coagulometer (Diagnostica Stago; Parsippany, New Jersey). A log-log plot of clotting times against dilutions of control pooled plasma in saline between 1:10 and 1:80 yielded linear standard reference curves. The clotting time of the 1:10 dilution of the control pooled plasma was assigned a value of 100% P&P activity, and clotting times of test samples were converted into percent of the control pooled plasma activity.
Specific Prothrombin Assay (Factor II)
![]()
Prothrombin activity was assayed by a one-stage assay in which a mixture of 100 µL of a prothrombin-depleted human serum/barium-adsorbed bovine plasma reagent [8] and 100 µL of a 1:10 to 1:40 dilution of test plasma were clotted by the addition of 200 µL of a thromboplastin C reagent containing CaCl2. Clotting times were converted to percent normal plasma prothrombin activity from a log-log standard curve prepared with dilutions of control pooled plasma.
Native Prothrombin Antigen Assay
![]()
Plasma native prothrombin antigen concentration was measured with native prothrombin antigen enzyme immunoassay kits provided by Organon Teknika. Color was measured at 450 nm with a Thermomax enzyme-linked immunosorbent assay reader (Molecular Devices; Menlo Park, California).
Results
![]()
Top
Methods
Results
Discussion
Author & Article Info
References
Relations between the Prothrombin Time Ratio and International Normalized Ratio Values Obtained with Six Thromboplastins and with the Coumatrak Monitor
|
|
An INR range between 2.0 and 3.5 has been recommended as the therapeutic range for patients receiving warfarin in the United States [9], whereas a higher range with upper limits between 4.5 and 4.8 are used in some European countries [10, 11]. Table 1 shows the relation of selected INR values obtained with Dade IS, with the other thromboplastins, and with the Coumatrak monitor as calculated from the linear regression lines of Figure 2 B. At a Dade IS INR of 2.0, the INRs for the plasma prothrombin times obtained with the other thromboplastins were all also essentially 2.0, whereas the INR obtained with the Coumatrak monitor was 2.3. At the upper limit of recommended therapeutic ranges, the Coumatrak monitor INRs corresponded closely to the INRs obtained from plasma prothrombin times done with Dade IS. In contrast, the INRs obtained with three thromboplastins with high ISI values substantially exceeded recommended upper limits as defined with Dade IS thromboplastin.
|
Relation between Prothrombin-Proconvertin Percent Activity and International Normalized Ratio Values
|
|---|
|
|
Because determinations of residual specific prothrombin activity [14] and native prothrombin antigen [7, 15, 16] have been proposed as being better techniques for monitoring oral anticoagulant therapy than the prothrombin time [14, 15], we also examined the relation between these measurements and INR values. We confirmed an earlier report of a good correlation between residual plasma native prothrombin antigen levels as measured by enzyme immunoassay and residual specific prothrombin activity as measured by a one-stage coagulation method [17] (r = 0.92, n = 89; data not shown).
The reciprocal of Dade IS INRs is plotted against plasma native prothrombin antigen levels in Figure 4 A and against plasma prothrombin activity in Figure 5 A. The linear regression lines from similar plots with each thromboplastin (excluding Coumatrak) are plotted in Figures 4B and 5B. The range of correlation coefficients for these regression lines were as follows: from 0.66 to 0.83 for native prothrombin antigen and from 0.76 to 0.88 for specific prothrombin activity. Table 2 summarizes, as calculated from Figures 4B and 5B, the relation between selected mean INR values and residual native prothrombin measured as antigen and activity. A mean INR range of 2.0 to 3.0 corresponded to between about 40% to 20% residual native plasma prothrombin.
|
|
Discussion
|
|---|
|
|
|---|
The variation in INR values for the thromboplastins with lower sensitivity (and therefore the higher ISI values) at least partly reflects the exponential function of the ISI in calculating an INR, which magnifies errors derived from any source as the ISI value increases. In contrast, two high-sensitivity thromboplastins, with ISI values approaching 1.0, gave virtually identical INR values Figure 2 Band Table 1.
It was somewhat reassuring that the plasma prothrombin time INRs obtained with all thromboplastins were acceptable in the lower portion of a recently recommended therapeutic INR range of 2.0 to 3.5 [9]. Unfortunately, the same was not true for values in the higher portion of the therapeutic range. At an INR of 3.5 obtained with the thromboplastin with the highest sensitivity (ISI, 1.27), three low-sensitivity thromboplastins yielded INRs exceeding this upper limit of the therapeutic range (Table 1). At an INR of 4.5 obtained with the highest-sensitivity thromboplastin, which is an INR within some accepted European therapeutic ranges for arterial thromboembolic disease [10, 11], the same three low-sensitivity thromboplastins yielded INRs that would definitely have led clinicians to reduce the dosage of warfarin prescribed (Table 1).
Thus, our data provide additional support for the recommendation [23] that all laboratories use thromboplastins with high sensitivity (low ISI values) for doing prothrombin times. Until this recommendation is widely adopted, we believe that clinicians monitoring therapy with plasma prothrombin times should, whenever possible, refer patients to a laboratory using a low ISI thromboplastin. This suggestion is particularly important for those patients in whom antithrombotic protection requires keeping the INR at the upper limit of a therapeutic range.
Capillary blood prothrombin time monitors are being increasingly used in outpatient anticoagulation clinics, whereas plasma prothrombin times are virtually always used to monitor therapy in hospitalized patients or patients being followed by private physicians. In our study, the INRs obtained on capillary blood correlated well with the INRs obtained from plasma prothrombin times done with high-sensitivity thromboplastins except for a positive bias at the lower limit of the therapeutic range Figure 2 Band Table 1, which others have also noted [24, 25]. As long as the clinician is aware of this positive bias, serious problems should not arise from reporting results as INR values within an institution using the Coumatrak capillary prothrombin time monitor for outpatients and a plasma prothrombin time with a high sensitivity thromboplastin for inpatients.
The P&P test was introduced in Scandinavia in the 1950s [6] to monitor oral anticoagulant therapy with a recommended therapeutic range of 10% to 30% [12]. The P&P test may be valuable in monitoring anticoagulant therapy in patients with a strong lupus anticoagulant prolonging a patient's baseline prothrombin time. Because the INR is determined from a prothrombin time ratio obtained by dividing the patient's prothrombin time on warfarin by a normal reference prothrombin time, not by the patient's baseline prothrombin time, the patient's INR will reflect a combined effect of warfarin and the underlying lupus anticoagulant on the patient's prothrombin time. Thus, the use of the usual INR therapeutic range could result in inadequate anticoagulation.
The percent P&P activity is a better measure of the effect of warfarin on the vitamin K-dependent factors affecting the prothrombin time in a patient with a strong lupus anticoagulant, because the 1/10 dilution of the test sample in the P&P test eliminates the inhibitory effect of a lupus anticoagulant upon the clotting time. As done in our study with the commercial reagent available in the United States in 1992, an 11% to 30% P&P range corresponded to mean INR values of 2.0 to 3.5 (Table 2). Thus, relating a P&P test result in a patient with lupus anticoagulant to the patient's corresponding INR value should allow the clinician to select a target INR for subsequent monitoring.
Hemorrhage and thrombosis in patients receiving an oral anticoagulant have been reported to be more closely related to specific plasma prothrombin activity [14] and to residual native prothrombin antigen [15] than to the prothrombin time test result. As first described by others [16, 17], native prothrombin antigen levels and specific plasma prothrombin activity were closely correlated. In our study, an INR range of 2.0 to 3.5 corresponded to a reduced normal plasma prothrombin content of about 40% to 15% of our normal reference pool when measured both as native prothrombin antigen and as specific residual prothrombin activity (Table 2).
Recent in vivo [8] and earlier in vitro [26] experiments provide convincing evidence that the antithrombotic effect of warfarin is strongly dependent on a reduction of plasma prothrombin activity. Therefore, we interpret our data to mean that, under most circumstances, reducing plasma prothrombin activity to 20% to 40% of normal is both required and sufficient to support the antithrombotic efficacy of warfarin. This information should be of value to a physician who may wish to confirm, by a specific prothrombin activity assay, the adequacy of oral anticoagulant therapy in particular situations as, for example, in deciding when after starting warfarin it is safe to withdraw heparin in a patient with extensive venous thrombotic disease.
Our data add to the evidence that the clinician may not yet assume that the use of the INR normalizes prothrombin time results obtained with different thromboplastins. A high-sensitivity thromboplastin with a low ISI value will give a more reliable INR, and, whenever possible, clinicians monitoring patients with plasma prothrombin times should refer patients to laboratories using such a thromboplastin.
Author and Article Information
|
|---|
|
|
|---|
References
|
|---|
|
|
|---|
1. ICSH/ICTH recommendations for reporting prothrombin time in oral anticoagulant control. International Committee for Standardization in Haematology and International Committee on Thrombosis and Haemostasis. J Clin Pathol. 1985; 38:133-4.
2. Hirsh J. Is the dose of warfarin prescribed by American physicians unnecessarily high? Arch Intern Med. 1987; 147:769-71.
3. Bussey HI, Force RW, Bianco TM, Leonard AD. Reliance on prothrombin time ratios causes significant errors in anticoagulation therapy. Arch Intern Med. 1992; 152:278-82.
4. Hirsh J. Substandard monitoring of warfarin in North America. Time for change (Editorial). Arch Intern Med. 1992; 152:257-8.
5. WHO Expert Committee on Biological Standardization. Technical Report Series 687, 33d Report. Geneva: World Health Organization; 1983:81-105.
6. Owren PA, Aas K. Control of Dicumarol therapy and quantitative determination of prothrombin and proconvertin. Scand J Clin Lab Invest. 1951; 3:201-8.
7. Furie B, Liebman HA, Blanchard RA, Coleman MS, Kruger SF, Furie BC. Comparison of the native prothrombin antigen and the prothrombin time for monitoring oral anticoagulant therapy. Blood. 1984; 64:445-51.
8. Zivelin A, Rao LV, Rapaport SI. Mechanism of the anticoagulant effect of warfarin as evaluated in rabbits by selective depression of individual procoagulant vitamin K-dependent clotting factors. J Clin Invest. 1993; 92:2131-40.
9. Hirsh J, Dalen JE, Deykin D, Poller L. Oral anticoagulants. Mechanism of action, clinical effectiveness, and optimal therapeutic range. Chest. 1992; 102:312S-326S.
10. Loeliger EA. Therapeutic target values in oral anticoagulationjustification of Dutch policy and a warning against the so-called moderate-intensity regimens. Ann Hematol. 1992; 64:60-5.
11. Smith P, Arnesen H, Holme I. The effect of warfarin on mortality and reinfarction after myocardial infarction. N Engl J Med. 1990; 323:147-52.
12. Owren PA. Long-term Dicumarol treatment in cardiovascular diseasetechnique and results. In: International Conference on Thrombosis and Embolism, 1st, Basel, 1954. Basel: Schwabe; 1955: 1085-94.
13. Bjerklund CJ. Effect of long term treatment with dicumarol in myocardial infarction. Thesis. Oslo University, Oslo, Norway: Oslo University Press; 1957.
14. Sise HS, Lavelle SM, Adamis D, Becker R. Relation of hemorrhage and thrombosis to prothrombin during treatment with coumarin-type anticoagulants. N Engl J Med. 1958; 259:266-71.
15. Furie B, Diuguid CF, Jacobs M, Diuguid DL, Furie BC. Randomized prospective trial comparing the native prothrombin antigen with the prothrombin time for monitoring oral anticoagulation therapy. Blood. 1990; 75:344-9.
16. Braun PJ, Szewczyk KM. Relationship between total prothrombin, native prothrombin and the International Normalized Ratio (INR). Thromb Haemost. 1992; 68:160-4.
17. Blanchard RA, Furie BC, Kruger SF, Waneck G, Jorgensen MJ, Furie B. Immunoassays of human prothrombin species which correlate with functional coagulant activities. J Lab Clin Med. 1983; 101:242-55.
18. Poller L, Thomson JM, Taberner DA. Effect of automation on prothrombin time test in NEQAS surveys. J Clin Pathol. 1989; 42:97-100.
19. van Rijn JL, Schmidt NA, Rutten WP. Correction of instrument- and reagent-based differences in determination of the International Normalized Ratio (INR) for monitoring anticoagulant therapy. Clin Chem. 1989; 35:840-3.
20. Poggio M, van den Besselaar AM, van der Velde EA, Bertina RM. The effect of some instruments for prothrombin time testing on the International Sensitivity Index (ISI) of two rabbit tissue thromboplastin reagents. Thromb Haemost. 1989; 62:868-74.
21. Kazama M, Suzuki S, Abe T, Tahara C, Shimazu C, Akiyama Y, et al. Evaluation of International Normalized Ratios by a controlled field survey with 4 different thromboplastin reagents. Thromb Haemost. 1990; 64:535-41.
22. Ng VL, Levin J, Corash L, Gottfried EL. Failure of the International Normalized Ratio to generate consistent results within a local medical community. Am J Clin Pathol. 1993; 99:689-94.
23. Taberner DA, Poller L, Thomson JM, Darby KV. Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR). J Clin Pathol. 1989; 42:92-6.
24. McCurdy SA, White RH. Accuracy and precision of a portable anticoagulation monitor in a clinical setting. Arch Intern Med. 1992; 152:589-92.
25. Becker DM, Humphries JE, Walker FB 4th, DeMong LK, Bopp JS, Acker MN. Standardizing the prothrombin time. Calibrating coagulation instruments as well as thromboplastin. Arch Pathol Lab Med. 1993; 117:602-5.
26. Xi M, Beguin S, Hemker HC. The relative importance of the factors II, VII, IX and X for the prothrombinase activity in plasma of orally anticoagulated patients. Thromb Haemost. 1989; 62:788-91.
This article has been cited by other articles:
![]() |
J. Ansell, J. Hirsh, E. Hylek, A. Jacobson, M. Crowther, and G. Palareti Pharmacology and Management of the Vitamin K Antagonists: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines (8th Edition) Chest, June 1, 2008; 133(6_suppl): 160S - 198S. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Wittkowsky Anticoagulation in Hypercoagulable Conditions: Special Considerations in the Treatment and Prevention of Venous Thromboembolism Journal of Pharmacy Practice, October 1, 2004; 17(5): 308 - 316. [Abstract] [PDF] |
||||
![]() |
J. Ansell, J. Hirsh, L. Poller, H. Bussey, A. Jacobson, and E. Hylek The Pharmacology and Management of the Vitamin K Antagonists: The Seventh ACCP Conference on Antithrombotic and Thrombolytic Therapy Chest, September 1, 2004; 126(3_suppl): 204S - 233S. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S Trask, R. C Gosselin, J. A Diaz, and W. E Dager Warfarin Initiation and Monitoring with Clotting Factors II, VII, and X Ann. Pharmacother., February 1, 2004; 38(2): 251 - 256. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hirsh, V. Fuster, J. Ansell, and J. L. Halperin American Heart Association/American College of Cardiology Foundation guide to warfarin therapy J. Am. Coll. Cardiol., May 7, 2003; 41(9): 1633 - 1652. [Full Text] [PDF] |
||||
![]() |
J. Hirsh, V. Fuster, J. Ansell, and J. L. Halperin American Heart Association/American College of Cardiology Foundation Guide to Warfarin Therapy Circulation, April 1, 2003; 107(12): 1692 - 1711. [Full Text] [PDF] |
||||
![]() |
J. Hirsh, J. E. Dalen, D. R. Anderson, L. Poller, H. Bussey, J. Ansell, and D. Deykin Oral Anticoagulants: Mechanism of Action, Clinical Effectiveness, and Optimal Therapeutic Range Chest, January 1, 2001; 119(1_suppl): 8S - 21S. [Full Text] [PDF] |
||||
![]() |
J. Ansell, J. Hirsh, J. Dalen, H. Bussey, D. Anderson, L. Poller, A. Jacobson, D. Deykin, and D. Matchar Managing Oral Anticoagulant Therapy Chest, January 1, 2001; 119(1_suppl): 22S - 38S. [Full Text] [PDF] |
||||
![]() |
S. S. Anand and S. Yusuf Oral Anticoagulant Therapy in Patients With Coronary Artery Disease: A Meta-analysis JAMA, December 1, 1999; 282(21): 2058 - 2067. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Wahl Dental Surgery in Anticoagulated Patients Arch Intern Med, August 10, 1998; 158(15): 1610 - 1616. [Abstract] [Full Text] |
||||
![]() |
T. R. Lousberg, D. M. Witt, D. G. Beall, B. L. Carter, and D. C. Malone Evaluation of Excessive Anticoagulation in a Group Model Health Maintenance Organization Arch Intern Med, March 9, 1998; 158(5): 528 - 534. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. M. Feinberg, E. S. Cornell, S. D. Nightingale, L. A. Pearce, R. P. Tracy, R. G. Hart, and E. G. Bovill Relationship Between Prothrombin Activation Fragment F1.2 and International Normalized Ratio in Patients With Atrial Fibrillation Stroke, June 1, 1997; 28(6): 1101 - 1106. [Abstract] [Full Text] |
||||
![]() |
F. Violi, D. Ferro, G. Valesini, S. I. Rapaport, D. T. Le, A. Slivka, E. Walz, A. J. Cole, S. Z. Nasr, A. L. Parke, et al. Thrombosis in the Antiphospholipid-Antibody Syndrome N. Engl. J. Med., September 7, 1995; 333(10): 665 - 667. [Full Text] |
||||
![]() |
L. B. Goldstein, A. J. Bonito, D. B. Matchar, P. W. Duncan, G. H. DeFriese, E. Z. Oddone, J. E. Paul, D. R. Akin, and G. P. Samsa US National Survey of Physician Practices for the Secondary and Tertiary Prevention of Ischemic Stroke : Design, Service Availability, and Common Practices Stroke, September 1, 1995; 26(9): 1607 - 1615. [Abstract] [Full Text] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||